| - | ====== CI ====== | + | ===== CI ===== |
| + | === Lamp Scan === | ||
| - | This routine records a scan of the lamp port based onthe Brewer model: | + | CI is used to perform a wavelength scan from 286.5 nm to 366 nm on one of the two internal test lamps. Filter Wheels are set to positions 1 and 0, and the Iris is open. The increment of the scan is user selectable with choices of 1, 2, or 5 Angstrom. Data can be compared with the data contained in the Final Test Record, and is stored in the file CIJJJYY.NNN in the instrument \DATA directory. |
| + | |||
| + | This routine records a scan of the lamp port based on the Brewer model: | ||
| ^Model ^ Wavelengths ^ | ^Model ^ Wavelengths ^ | ||
| |MKII|286.5 to 325 nm| | |MKII|286.5 to 325 nm| | ||
ci.1605276471.txt.gz · Last modified: 2020/11/16 13:55 (external edit)