-====== CI ======+===== CI ===== 
 +=== Lamp Scan ===
  
-This routine records a scan of the lamp port based onthe Brewer model:+CI is used to perform a wavelength scan from 286.5 nm to 366 nm on one of the two internal test lamps. Filter Wheels are set to positions 1 and 0, and the Iris is open. The increment of the scan is user selectable with choices of 1, 2, or 5 Angstrom. Data can be compared with the data contained in the Final Test Record, and is stored in the file CIJJJYY.NNN in the instrument \DATA directory.   
 + 
 +This routine records a scan of the lamp port based on the Brewer model:
 ^Model ^ Wavelengths ^ ^Model ^ Wavelengths ^
 |MKII|286.5 to 325 nm| |MKII|286.5 to 325 nm|