| ====== HP ====== | ====== HP ====== | ||
| + | === Grating Synchronization === | ||
| - | HP is a routine that moves the bottom micrometer to find the location with the highest | + | HP is a routine that moves the bottom micrometer to optimize |
| - | Typically this routine is combined with FR and [[hg|HG]] | + | In double spectrometer instruments, |
| + | For this test the Standard lamp is used as a light source. The Standard Lamp is turned on, the prism is pointed to lamps, the Iris is opened, FW#1 is set to Position 1, and FW#2 is set to position 0. Micrometer #2 is then moved relative to Micrometer #1, and is adjusted such that maximum intensity will occur - an adjustment of more than 10 steps results in the test being repeated. | ||
| + | |||
| + | |||
| + | Typically this routine is combined with [[fr|FR]] and [[hg|HG]] | ||