====== Filter Wheel Setup and Uses ====== This is just the standard filter wheel setup used on the Canadian Brewer Spectrophotometer Network. Other networks may differ | ^ Filter Wheel 1 ^^^^ Filter Wheel 2 ^^^ ^Position|Steps|Attenuation|Filter|Notes|Steps|Filter|Notes| ^0|320|FP|Film Polarizer| 0|0|Clear| ^1|256|2.0|GQD2|Used for most measurements|64|0.5|ND1| ^2|192|3.0|ND6|Not used (Future Autosighting?)|128|1|ND2| ^3|128|0|Clear|Used during FZ, FM|192|1.5|ND3| ^4|64|4.5|GDQ2+2.5|Not Used|256|2.0|ND4| ^5|0|1.7|GDQ1|Used for DT/RS Measurement*|320|2.5|ND5| In order to use Position 5 in filter wheel one in DT/RS, this must be enabled in [[ic|IC]] Enter IC, and go down to "Second film polarizer" and press [key]h[/key]. This will change the name to "FW1@5 for DTRS" The Canadian Brewer Spectrophotometer Network has done this achieve higher counts during DT/RS test. The ideal counts for DT is 3 million counts on the high intensity and 1 million counts on the low intensity. For RS we are aiming for 1 million counts for best results. Shown below is the expected DI for filter wheel 1\\ {{:misc:fw1di.png?600|}}