===== MM ===== === TeleType Communications === MM sets the Brewer program to its teletype (terminal) mode, whereby the operator may communicate directly with the Brewer via the computer keyboard and a set of low level commands. \\ Exit from teletype mode by typing EXIT. See Appendix H for more explanation on the TeleType feature.\\ ===== List of commands ===== ^Command^Variables^Usage^Description^ |A| | |Repeats the command. Usually use in conjunction with another command move. (I.E. R,0,7,5:O:A) Press home to delete twice to go exit.| |B,x| \\ x refers to the following 0. Turns off all lamp 1. Turns on the mercury lamp 2. Turns on the standard lamp |B,2|This command controls the lamps| |M,m,s|\\ m refers to a motor number 1. Zenith prism 2. Azimuth Tracker 3. Iris 4. Filterwheel 1 5. Filterwheel 2 6. Filterwheel 3 9. Micrometer 2 (Bottom) 10. Micrometer 1 (Top) 11. Slitmask 1 12. Slitmask 2 13. Zenith tracker s refer to the number of steps to move |M,10,500|Moves motor m to step s. Note: when moving backwards (negative number) it will set that as the new reference point. [[motor|Click here more motor information]].| |O| |O|This command outputs the number of counts returned in a measurement. |R,p1,p2,t|\\ p# refers to the slit positions t refers to number of times to repeat |R,0,5,7|This command is used to measure the light intensity according to the variables setup. Often this command is combined with the O command to output the results. R,0,5,7:O| |SL| |SL| \\ - Turn the SL bulb on - Points the zenith down - Open the iris - Set FW1 to position 1 - Set FW2 to position 0 | |SL+| |SL| \\ - Turn the SL bulb on - Points the zenith down - Open the iris - Set FW1 to position 1 - Set FW2 to position 0 - R,0,7,1;o | |HG+| |HG+| \\ - Turn the SL bulb on - Points the zenith down - Open the iris - Set FW1 to position 1 - Set FW2 to position 0 - R,0,7,1;o | |FWfilter wheel number,position|filter wheel number Filter wheel 1 or 2 \\ position - What position to move to|FW1,2|Move the selected filter wheel to the selected position| |TStart Location2End Location| \\ Start Location and End Location can be any of the following 1. MAX - Maximum micrometer position 2. OP - operating micrometer position 3. MIN - minimum micrometer position |TOP2MAX|This command will move the micrometer 1 to max/op/min position to max/op/min. It assumes the micrometer is already in the start location. This is use to help test the micrometer range.| |BStart Location2End Location| \\ Start Location and End Location can be any of the following 1. MAX - Maximum micrometer position 2. OP - operating micrometer position 3. MIN - minimum micrometer position |BOP2MAX|This command will move the micrometer 2 to max/op/min position to max/op/min. It assumes the micrometer is already in the start location. This is use to help test the micrometer range.| |exit| |exit|exit's mm and return to menu ('quit' will do the same thing)| \\ ==== For multiboard electronics only ==== ^Command^Variables^Usage^Description^ |E,m| \\ m refers to a motor number 1. Zenith prism 2. Azimuth Tracker 3. Iris 4. Filterwheel 1 5. Filterwheel 2 6. Filterwheel 3 or Micrometer 2 (bottom) 10. Micrometer 1 (Top) 11. Slitmask 1 12. Slitmask 2 13. Zenith tracker |E,9|This command zeros the motor and return the number it takes to get to the LED. NOTE: This will not return the motor and will leave it at the LED| |x,s1,s2|s1 and s2 refers to the number of steps to move for micrometer 1 and 2 respectively.|x,100,200|Moves both micrometer 1 and 2 to step s. Note: when moving backwards (negative number) it will set that as the new reference point. [[motor|Click here more motor information]].| ==== For new electronics only ==== ^Command^Variables^Usage^Description^ |? OR !|See [[operation_information|Operation Information]]|:::|A variety of operational commands/status output for the brewer. for more.| |I,m| \\ m refers to a motor number 1. Zenith prism 2. Azimuth Tracker 3. Iris 4. Filterwheel 1 5. Filterwheel 2 6. Filterwheel 3 9. Micrometer 2 (Bottom) 10. Micrometer 1 (Top) 11. Slitmask 1 12. Slitmask 2 13. Zenith tracker | |I,9|This commands re-initializes the specified motor| |N,s|s refers to the number of steps|N,500|(Used for [[singleboard|singleboard electronics]] only.) Moves both micrometer 1 and 2 to step s. Note: when moving backwards (negative number) it will set that as the new reference point. [[motor|Click here more motor information]].| |SAVE| |SAVE|Stores the current set of RAM configuration parameters| |?MZ10| |?MZ10|Runs the command ?MOTOR.ZERO.POS[10]| |?MZ9| |?MZ9|Runs the command ?MOTOR.ZERO.POS[9]| |?MP10| |?MP10|Runs the command ?MOTOR.POS[10]| |?MP9| |?MP9|Runs the command ?MOTOR.POS[9]| |?MD10| |?MD10|Runs the command ?MOTOR.DISCREPANCY[10]| |?MD9| |?MD9|Runs the command ?MOTOR.DISCREPANCY[10]|