====== All about Spectrometer ====== This page is mainly split to Three different section. - Common information - Single Spectrophotometer (MKII, and MKIV) - Double Spectrophotometer (MKIII) ===== Common Information ===== ==== Direction of the lens ==== Object should appear tall and thin and concave surface should be facing (Direction of light travel) towards the slits or PMT\\ {{:pictures:correction_lens.png?400|}} \\ Correction Lens The correction lens [LE 5] has a convex-cylindrical surface (radius 170.0 mm) and a concave-spherical surface (radius 230.0 mm). Both surfaces are coated with a layer of magnesium fluoride to minimize reflectance at 315.0 nm for an incidence angle of 29°. The lens is mounted in the entrance-slit housing at an angle of 29° to the optical axis with the concave-spherical surface facing the entrance slit. The axis of the cylindrical surface is positioned in the horizontal plane to within 1°. ===== Single Spectrophotometer ===== ===== Double Spectrophotometer ===== ==== Disassembly ==== To remove the Bulkhead from the spectrometer, there are 4 screws that needs to be removed. *NOTE* Make sure the bulkhead is being held onto before taking all the screws off!\\ {{:pictures:20231122_091909.jpg?400|}} {{:pictures:20231122_111902.jpg?400|}}