-====== CI ======+===== CI ===== 
 +=== Lamp Scan ===
  
-This routine records a scan of the lamp port based onthe Brewer model:+CI is used to perform a wavelength scan from 286.5 nm to 366 nm on one of the two internal test lamps. Filter Wheels are set to positions 1 and 0, and the Iris is open. The increment of the scan is user selectable with choices of 1, 2, or 5 Angstrom. Data can be compared with the data contained in the Final Test Record, and is stored in the file CIJJJYY.NNN in the instrument \DATA directory.   
 + 
 +This routine records a scan of the lamp port based on the Brewer model:
 ^Model ^ Wavelengths ^ ^Model ^ Wavelengths ^
 |MKII|286.5 to 325 nm| |MKII|286.5 to 325 nm|
Line 16: Line 19:
  
 Uses: 2090, 2450, 3220, 3230, 4450, 6610, 6630, 6650, 6700, 7000, 8350, 9450, 9500, 9650, 9805, 9840 Uses: 2090, 2450, 3220, 3230, 4450, 6610, 6630, 6650, 6700, 7000, 8350, 9450, 9500, 9650, 9805, 9840
 +
 +This Routine will only measure on one direction, for measuring both up and down the wavelength, use [[CJ|CJ]]
  
  