| ====== Filter Wheel Setup and Uses ====== | ====== Filter Wheel Setup and Uses ====== | ||
| - | This is just the standard filter wheel setup used on the Canadian Spectrophotometer Network. Other countries | + | This is just the standard filter wheel setup used on the Canadian |
| - | ^< | + | | ^ |
| - | |Position|Attenuation|Filter|Notes|Position|Filter|Notes| | + | ^Position|Steps|Attenuation|Filter|Notes|Steps|Filter|Notes| |
| - | |0|FP|Film Polarizer| |0|0|Clear| | + | ^0|320|FP|Film Polarizer| 0|0|Clear| |
| - | |1|2.0|GQD2|Used for most measurements|1|0.5|ND1| | + | ^1|256|2.0|GQD2|Used for most measurements|64|0.5|ND1| |
| - | |2|3.0|ND6|Not used (Future Autosighting? | + | ^2|192|3.0|ND6|Not used (Future Autosighting? |
| - | |3|0|Clear|Used during FZ, FM|3|1.5|ND3| | + | ^3|128|0|Clear|Used during FZ, FM|192|1.5|ND3| |
| - | |4|4.5|GDQ2+2.5|Not Used|4|2.0|ND4| | + | ^4|64|4.5|GDQ2+2.5|Not Used|256|2.0|ND4| |
| - | |5|1.7|GDQ1|Used for DT/RS Measurement*|5|2.5|ND5| | + | ^5|0|1.7|GDQ1|Used for DT/RS Measurement*|320|2.5|ND5| |
| + | In order to use Position 5 in filter wheel one in DT/RS, this must be enabled in [[ic|IC]] | ||
| + | Enter IC, and go down to " | ||
| + | The Canadian Brewer Spectrophotometer Network has done this achieve higher counts during DT/RS test. The ideal counts for DT is 3 million counts on the high intensity and 1 million counts on the low intensity. For RS we are aiming for 1 million counts for best results. | ||
| + | Shown below is the expected DI for filter wheel 1\\ | ||
| + | {{: | ||