 ===== FR ===== ===== FR =====
-=== Micrometer Reset === +==== Micrometer Reset ==== 
 FR performs a reset of the wavelength adjust micrometers. The micrometers are moved to reference points, and then moved to the operating points as dictated by offset values contained in the instruments constants file.\\ FR performs a reset of the wavelength adjust micrometers. The micrometers are moved to reference points, and then moved to the operating points as dictated by offset values contained in the instruments constants file.\\
    
Line 6: Line 6:
  
 FR results are recorded in the average file, MIOAVG.nnn. FR results are recorded in the average file, MIOAVG.nnn.
 +The layout for the MIOAVG file is as follows
 +
 +^Julian Date^Measuring mode^Temperature^Cal step^Top Micrometer found at^Top Micrometer set at^Bottom Micrometer found at^Bottom Micrometer set at^ 
 +|  08525  |  o3  |  8  |  1055  |  7173  |  7301  |  6101  |  6094  |
 +==== Associated commands ====
 +=== FT ===
 +The FT command performs a Find Reference FR command but only for the top micrometer in double spectrometer instruments. (Will work with a single spectrometer instrument, but will react the same as a FR command)
 +=== FB ===
 +The FB command performs a Find Reference FR command but only on the bottom micrometer for double spectrometer instruments.
 +
 +
 +
 +