hh
Differences
This shows you the differences between two versions of the page.
| Next revision | Previous revision | ||
| hh [2020/12/02 16:37] – created reno | hh [2021/11/25 00:18] (current) – mbrohart | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| ====== HH ====== | ====== HH ====== | ||
| + | === Micrometer Positioning test === | ||
| - | HH is will run an [[hg|HG]] except it ignores | + | This routine |
| + | |||
| + | The test can indicate the accuracy of the dispersion and the accuracy of the spherical mirror placement within the instrument by measuring the wavelength of the emission line/s against the ideal wavelength and recording the measured slit width (FHHB - Full height, Half Base of the scanned emission line image). | ||
hh.1606927035.txt.gz · Last modified: 2020/12/02 16:37 by reno