AP
Monitor Voltages Printout
This command prints to the monitor screen, or to disc, a number of diagnostics that are continuously available in the Brewer. The diagnostics include power supply voltages, test lamp voltages and currents, temperatures, and Brewer moisture content.
Typical AP Command Results:
Monitoring specific lines
You are able to monitor only specific lines in ap with the following command.
Ap@15 - This will only monitor line 15 on AP while lamps is on. Change 15 to any other number to monitor other lines. Ap@15_16 - This will monitor line 15 and 16 on AP while lamps is on. The _ can be treated like the &. You can add more lines by adding more numbers like Ap@1_2_3_4 and etc.