| - | ===== HVB ===== | + | ===== HVB (Depreciated)===== |
| + | NOTE: As of version 7.33, HVB has been depreciated. [[hv|HV]] will be used for both multiboard & single board electronics. | ||
| === High Voltage Test (MB) === | === High Voltage Test (MB) === | ||
| The HVB command | The HVB command | ||
| Line 5: | Line 6: | ||
| The HV command invokes a test used to determine the optimum high voltage setting for the photomultiplier. \\ | The HV command invokes a test used to determine the optimum high voltage setting for the photomultiplier. \\ | ||
| - | This test is normally used in the factory during the final set-up stages of manufacture, | ||
| The Standard Lamp is turned on, the iris is opened, FW#1 is put to position 1, and FW#2 is set to position 0 | The Standard Lamp is turned on, the iris is opened, FW#1 is put to position 1, and FW#2 is set to position 0 | ||
| To run an automatic high voltage test for single board (new electronics) use [[hv|HV]]. \\ | To run an automatic high voltage test for single board (new electronics) use [[hv|HV]]. \\ | ||
| - | Procedural steps to prepare | + | === Procedure (MB) === |
| - | + | - Ensure the instrument is able to run diagnostic data (As it still required to measure the SL Bulb) | |
| + | - Turn the instrument | ||
| + | - Take the lid off the Brewer. | ||
| + | - Take the cover off the Secondary power supply | ||
| + | - Turn the brewer on | ||
| + | - Run the command " | ||
| + | - Follow the dosbox prompt (See Suggested Values | ||
| + | - During the lamp warm up, set the multimeter to the voltage setting, and insert the leads to GND and HV on the panel beside the secondary power supply | ||
| + | - Once the lamp has warmed up and is waiting for the high voltage to be set, adjust the R202 pot until you are at the correct voltage - Refer to the table below for more info. Press the top left sighting button | ||
| + | - The Iris will move to the from the middle to the open position, then back to the middle, Once it is at the middle you may proceed to next voltage | ||
| + | - Repeat until you have completed the High voltage test | ||
| + | === Suggested values === | ||
| + | ^ ^Description^Suggested values^ | ||
| + | |Photomultiplier serial Number: | ||
| + | |Preame Voltage level|Enter the discriminator (PMT) voltage|52mm PMT are typically around -30mV, where as 22mm PMT are usually around -10mV| | ||
| + | |Min Voltage|Where to start measuring the high voltage test from|For MB, usually around -900~-1000v| | ||
| + | |Max Voltage|Where to end the high voltage test at|For MV, usually around -1700| | ||
| + | |Increment|Increment per level|For MB, 50, for SB, 25| | ||
| + | |||
| + | ^High Voltage(V)^ Measured Voltage(V)^ | ||
| + | |900|-4.5| | ||
| + | |950|-4.75| | ||
| + | |1000|-5.00| | ||
| + | |1050|-5.20| | ||
| + | |1100|-5.50| | ||
| + | |1150|-5.75| | ||
| + | |1200|-6.00| | ||
| + | |1250|-6.25| | ||
| + | |1300|-6.50| | ||
| + | |1350|-6.75| | ||
| + | |1400|-7.00| | ||
| + | |1450|-7.25| | ||
| + | |1500|-7.50| | ||
| + | |1550|-7.75| | ||
| + | |1600|-8.00| | ||
| + | |1650|-8.25| | ||
| + | |1700|-8.50| | ||
| + | |1750|-8.75| | ||
| + | |1800|-9.00| | ||