highspeedamp
High Speed Amp
Adjusting and Measuring the Discriminator Voltage
The default setting for the discriminator voltage depends on what type of PMT is in use. For 28mm PMT (New PMT or in most MKIII/New electronics) the typically setting is around -10mV. for 52mm PMT, the typically setting is around -30mV.
To adjust, measure from TP1 and the ground on the BNC connector shell and adjust R1.