sl
SL
Standard Lamp Test
SL initiates the most important quality assurance test in the Brewer commend set. This test essentially performs an ozone measurement using an internal quartz-halogen lamp as the source. The test should be run at least at the start and end of the day and should be preceded by an hp and an hg, The values of R5, R6, and F1 should be monitored carefully, and any changes should be noted and investigated.
For the SL test, the iris is opened, FW#1 is set to position 1, and FW#2 is set to position 0
Refer to Lamps
sl.txt · Last modified: 2021/11/25 01:56 by mbrohart