MM

MM is a command that puts you into the teletype (terminal) of the Brewer Spectrometer.

List of commands

CommandVariablesUsageDescription
B,x
x refers to the following
0. Turns off all lamp
1. Turns on the mercury lamp
2. Turns on the standard lamp
B,2This command controls the lamps
M,m,s
m refers to a motor number
1. Zenith prism
2. Azimuth Tracker
3. Iris
4. Filterwheel 1
5. Filterwheel 2
6. Filterwheel 3
9. Micrometer 2 (Bottom)
10. Micrometer 1 (Top)
11. Slitmask 1
12. Slitmask 2
13. Zenith tracker
M,10,500Moves motor m to step s. Note: when moving backwards (negative number) it will set that as the new reference point
O O
R,p1,p2,t
p# refers to the slit positions
t refers to number of times to repeat
R,0,5,7This command is used to measure the light intensity according to the variables setup. Often this command is combined with the O command to output the results. R,0,5,7:O

For new electronics only

CommandVariablesUsageDescription
I,m
m refers to a motor number
1. Zenith prism
2. Azimuth Tracker
3. Iris
4. Filterwheel 1
5. Filterwheel 2
6. Filterwheel 3
9. Micrometer 2 (Bottom)
10. Micrometer 1 (Top)
11. Slitmask 1
12. Slitmask 2
13. Zenith tracker
I,9This commands re-initializes the specified motor
SAVESAVEStores the current set of RAM configuration parameters
mm.1605810015.txt.gz · Last modified: 2020/11/19 18:20 by sitr