SHICCal
Setup
Prerequisite
- SHIC software must be reside in your root directory.
C:\SHICrivm
- UAView software
- All ### should be changed to the respective instrument number
Processing UV data
The UV data must first be process before SHIC is able to ingest the data.
- Place the UV file and the corresponding UVR file inside the same directory, IE:
C:\SHICrivm\uvdata\temp
- Run the UA Scan program and process one UV file at a time. Note: once you click process, you must highlight one scan, then press SaveScans
- Run the following .bat file to rename the processed files (removed _YY from the end)
- rename.bat
@@echo off setlocal enabledelayedexpansion set /p brewer="Enter Brewer #: " set "folder=C:\SHICrivm\uvdata\%brewer%" for %%a in ("%folder%\*_*.%brewer%") do ( set "oldname=%%~nxa" set "newname=!oldname:_=!" ren "%%a" "!newname!" )
- Move the .### files to
C:\SHICrivm\uvdata\###
- Inside C:\SHICrivm\uvanalys, create a file, ###.dfa with the following inside.
- ###.dfa
ozonshi.dfa 0 no_id_met .### c:\shicrivm\uvdata\###\ 285.00 440.00 0.50 1 0 0.550 1 1.0
- Launch SHICcall5_25.exe
- Set the instrument_ID to ### and the DAYtime to which ever you files you'd like to prcoess
- Click Run SHIC
**Note: there is a shicrivm.log inside the SHICCall directory that can help troubleshoot any issues
Functions
Spectral shifts - Helps determines any shift to the instrument, can help point to bad HG bulbs/DSP applied. Should be within +/- 0.1nm
Spectral spikes - shows if there is any issues with the instrument measuring. should be within +/- 10 percent after 310nm