MM
MM is a command that puts you into the teletype (terminal) of the Brewer Spectrometer.
List of commands
| Command | Variables | Usage | Description |
|---|---|---|---|
| A | Repeats the command. Usually use in conjunction with another command move. (I.E. R,0,7,5:O:A) Press home to delete twice to go exit. | ||
| B,x | x refers to the following 0. Turns off all lamp 1. Turns on the mercury lamp 2. Turns on the standard lamp | B,2 | This command controls the lamps |
| M,m,s | m refers to a motor number 1. Zenith prism 2. Azimuth Tracker 3. Iris 4. Filterwheel 1 5. Filterwheel 2 6. Filterwheel 3 9. Micrometer 2 (Bottom) 10. Micrometer 1 (Top) 11. Slitmask 1 12. Slitmask 2 13. Zenith tracker | M,10,500 | Moves motor m to step s. Note: when moving backwards (negative number) it will set that as the new reference point. Click here more motor information. |
| O | O | ||
| R,p1,p2,t | p# refers to the slit positions t refers to number of times to repeat | R,0,5,7 | This command is used to measure the light intensity according to the variables setup. Often this command is combined with the O command to output the results. R,0,5,7:O |
For new electronics only
| Command | Variables | Usage | Description |
|---|---|---|---|
| ?<command> OR !<command> | See Operation Information | A variety of operational commands/status output for the brewer. for more. | |
| I,m | m refers to a motor number 1. Zenith prism 2. Azimuth Tracker 3. Iris 4. Filterwheel 1 5. Filterwheel 2 6. Filterwheel 3 9. Micrometer 2 (Bottom) 10. Micrometer 1 (Top) 11. Slitmask 1 12. Slitmask 2 13. Zenith tracker | I,9 | This commands re-initializes the specified motor |
| SAVE | SAVE | Stores the current set of RAM configuration parameters | |